ACI Values for 34465A and 34470A DMMs
Input
Error from Nominal
Current, rms Frequency Range
Filter
Quick
Check
24 hours
90 days
1 year 2 years
1 mA
1 kHz
1 mA
200 Hz
Q
± 1.1 µA
± 1.3 µA
± 1.4 µA
5 kHz
100 µA
1 kHz
100 µA 200 Hz
± 0.110 µA ± 0.130 µA
± 0.140 µA
5 kHz
Q
:
Quick performance verification test point.
Performance Verification
520
Keysight True
volt
Series Operating and Service Guide