*OPC?
Returns 1 to the output buffer after all pending commands complete. Other commands cannot be
executed until this command completes.
Parameter
Typical Return
(none)
1
Configure 100 DC voltage measurements and return "1" when the measurements complete:
CONF:VOLT:DC
SAMP:COUN 100
INIT
*OPC?
l
Use this command to synchronize your application with the instrument.
l
The difference between *OPC and *OPC? is that*OPC sets a status bit when the operation completes,
and *OPC? outputs "1" when the operation completes.
*OPT?
Returns a string identifying any installed options.
Parameter Typical Return
(none)
(see below)
Return the installed options:
*OPT?
l
This query returns a string with the following format:
34460A: {GPB|0},{LAN|0},{SEC|0}
34461A: {GPB|0},{SEC|0}
34465A/70A: {GPB|0},{SEC|0},{DIG|0},{MEM|0}
A 0 is returned for options that are not installed. For example, a 34461A with only GPIB installed
returns GPB,0. A 34460A with only GPB and SEC installed returns GPB,0,SEC.
l
Options vary by model, as
SCPI Programming Reference
272
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...