[SENSe:]{RESistance|FRESistance}:RESolution {<
resolution
>|MIN|MAX|DEF}
[SENSe:]{RESistance|FRESistance}:RESolution? [{MIN|MAX|DEF}]
Selects the measurement resolution for all resistance measurements. Specify the resolution in the same
units as the selected measurement function, not in number of digits.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version
of this command or query is identical to using the RESistance version.
Parameter
Typical Return
<
resolution
>: See
or
The default
is equivalent to 10 PLC. Specify <
resolution
> in the measurement's units (V, A, Hz,
Ω, and so on).
+3.0000
Configure 2-wire resistance measurements with a 3 Ω resolution. Then make and read one measurement:
CONF:RES 1E6
RES:RES 3
READ?
Typical Response: +6.2705
l
You can specify MIN (best resolution) or MAX (worst resolution) instead of <
resolution
>.
l
To achieve normal mode (line frequency noise) rejection, use a resolution that corresponds to an integ-
ration time that is an integral number of power line cycles.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
See Also
[SENSe:]{RESistance|FRESistance}:NPLC
Keysight True
volt
Series Operating and Service Guide
355
SCPI Programming Reference
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...