Resolution and Integration Time for DC Measurements
Setting the integration time also sets the measurement resolution. This table shows the relationship
between integration time in power line cycles (PLC) and resolution.
Integration Time
Resolution
(34460A)
Resolution
(34461A)
Resolution
(34465A)
Resolution
(34470A)
Power Line
Cycles
Aperture
(60 Hz
Power)
Aperture
(50 Hz
Power)
0.001 PLC
1
20 µs
1
-
-
30 ppm x
Range
30 ppm x
Range
0.002 PLC
1
40 µs
1
-
-
15 ppm x
Range
10 ppm x
Range
0.006 PLC
1
100 µs
1
-
-
6 ppm x Range 3 ppm x Range
0.02 PLC
(MINimum)
0.3 ms
300 ppm × Range
(MAXimum)
100 ppm × Range
(MAXimum)
3 ppm x Range 1 ppm x Range
0.06 PLC
1 ms
-
-
1.5 ppm x
Range
0.5 ppm x
Range
0.2 PLC
3 ms
100 ppm × Range
10 ppm × Range
0.7 ppm x
Range
0.3 ppm x
Range
1 PLC
16.7 ms
20 ms
30 ppm × Range
3 ppm × Range
0.3 ppm x
Range
0.1 ppm x
Range
10 PLC
(DEFault)
0.167 s
0.2 s
10 ppm × Range
(DEFault)
1 ppm × Range
(DEFault)
0.1 ppm x
Range
0.03 ppm x
Range
100 PLC
(MAXimum)
1.67 s
2 s
3 ppm × Range
(MINimum)
0.3 ppm × Range
(MINimum)
0.03 ppm x
Range
0.01 ppm x
Range
1
34465A/70A with DIG option only.
Keysight True
volt
Series Operating and Service Guide
455
SCPI Programming Reference
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...