Capacitance Offset Calibration (Optional)
This calibration procedure applies to all models (34460A/61A/65A/70A).
This calibration is normally not required. It compensates for residual capacitance offset which is typically
less than a few pF. In normal use, any residual offset will be overridden by the null function (null also cor-
rects for cable capacitance), and the null function is required to achieve the multimeter’s capacitance spe-
cification.
Configuration: Capacitance
1.
Remove all connections from the instrument's front terminals.
2.
Select the front terminals.
3.
Enter a calibration value of +0 (see
). A Calibration Step Succeeded mes-
sage indicates success; if the display shows Calibration Step Failed, check the input value, range, func-
tion, and entered calibration value and repeat the calibration step.
4.
Repeat steps 1 through 3 for the instrument's rear terminals (not applicable to the 34460A).
5.
Store the new calibration constants (see
Storing the Calibration Constants
).
Keysight True
volt
Series Operating and Service Guide
569
Calibration Adjustment Procedures
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...