The following table lists the bit definitions for the
Standard Operation Register
:
Bit Num-
ber
Bit Name
Decimal
Value
Definition
0
Calibrating
1
Instrument is performing a calibration.
1
Not Used
2
(Reserved for future use)
2
Not Used
4
(Reserved for future use)
3
Not Used
8
(Reserved for future use)
4
Measuring
16
Instrument is initiated, and is making or about to make a measurement.
5
Waiting for
Trigger
32
Instrument is waiting for a trigger.
6
Not Used
64
(Reserved for future use)
7
Not Used
128
(Reserved for future use)
8
Configuration
Change
256
Instrument configuration has been changed since the last INIT, READ? or MEASure?, either from
the front panel or from SCPI.
9
Memory
Threshold
512
Programmed number of measurements (DATA:POINts:EVENt:THReshold) have been stored in
measurement memory.
10
Instrument
Locked
1024
Set if a remote interface (GPIB, USB or LAN) has a lock (SYSTem:LOCK:REQuest?). Cleared when
the remote interface releases the lock (SYSTem:LOCK:RELease).
11
Not Used
2048
(Reserved for future use)
12
Not Used
4096
(Reserved for future use)
13
Global Error
8192
Set if any remote interface has an error in its error queue; cleared otherwise.
14
Not Used
16384
(Reserved for future use)
15
Not Used
32768
"0" is returned.
Command Summary
STATus:QUEStionable:CONDition?
SCPI Programming Reference
388
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...