Power-On and Reset State
These tables show factory default settings. Parameters marked with a bullet (•) are nonvolatile and are not
affected by power-on or system reset. Other parameters are volatile and are reset to the indicated values
at power-on or after
or
.
The power-on/reset state may differ from these values if you have enabled
power-on state recall mode.
Factory Default Settings
Measurement Configuration
Factory Setting
Function
DC Volts
Range
Autorange (for all functions)
Integration Time
NPLC On, 10 power line cycles (PLC)*
Resolution
Resolution equivalent to 10 PLC
Autozero
On*
Aperture (34465A/70A only)*
Off, 100 ms
Frequency or Period Aperture (Gate Time)
100 ms*
• Input Impedance
• 10 MΩ (fixed for all DC voltage ranges)
AC Input Filter (bandwidth)
20 Hz (medium filter)
Temperature Transducer Type
FRTD
RTD Reference Resistance
100 Ω
Nulls (individual functions)
Off; reference = 0.0, auto reference enabled
(for all measurement functions)
Current Terminals
3 A
Secondary Measurements
Off
Screen Shot Format
PNG
Calibration Value
0.0
Current Switch Mode (34465A/70A only)
Continuous
Resistance (2- and 4-wire) (34465A/70A only):
Offset Compensation
Off
Low Power Measurements
Off
Temperature (34465A/70A only):
Offset Compensation (2- and 4- wire RTD)
Off
Low Power Measurements(2- and 4- wire RTD or thermistor) Off
Thermocouple (34465A/70A only):
Check
Off
Reference Junction Temperature, Type, Adjust
0° C, Internal, 0° C
Thermocouple Type
J
* For all DC measurement functions.
SCPI Programming Reference
474
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...