[SENSe:]VOLTage[:DC]:IMPedance:AUTO {ON|1|OFF|0}
[SENSe:]VOLTage[:DC]:IMPedance:AUTO?
Disables or enables automatic input impedance mode for DC voltage and ratio measurements.
Parameter
Typical Return
{ON|1|OFF|0}. Default: OFF.
0 (OFF) or 1 (ON)
Make all DC voltage measurements with a 10 MΩ input impedance:
VOLT:IMP:AUTO OFF
l
OFF:
the input impedance for DC voltage measurements is fixed at 10 MΩ for all ranges to minimize
noise pickup.
l
ON:
the input impedance for DC voltage measurements varies by range. It is set to "HI-Z" (>10 GΩ) for
the 100 mV, 1 V, and 10 V ranges to reduce the effects of measurement loading errors on these lower
ranges. The 100 V and 1000 V ranges remain at a 10 MΩ input impedance.
l
The
and
commands automatically select "AUTO OFF".
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
Keysight True
volt
Series Operating and Service Guide
381
SCPI Programming Reference