![Keysight 34460A Скачать руководство пользователя страница 279](http://html1.mh-extra.com/html/keysight/34460a/34460a_operating-and-service-manual_1961230279.webp)
MEASure Subsystem
The MEASure queries are the easiest way to program measurements because they always use default meas-
urement parameters. You set the function, range, and resolution in one command, but you cannot change
other parameters from their default values. The results are sent directly to the instrument's output buffer.
A MEASure query is functionally equivalent to sending CONFigure followed immediately by
. The
difference is that CONFigure commands allow you to change parameters between the CONFigure and the
READ?
Command Summary
MEASure:{RESistance|FRESistance}?
Default Settings for MEASure?
With the MEASure? queries, you can select the function, range and resolution in one command. All other
parameters are set to their default values (below).
Measurement Parameter
Default Setting
AC Input Filter (bandwidth)
20 Hz (medium filter)
Autozero
OFF if resolution setting results in NPLC < 1
ON if resolution setting results in NPLC ≥ 1
Range
AUTO (including voltage range for frequency and period measurements)
Samples per Trigger
1 sample
Trigger Count
1 trigger
Trigger Delay
Automatic delay
Trigger Source
Immediate
Trigger Slope
NEGative
Math Functions
Disabled. Other parameters are unchanged.
Per-function Null State
Disabled
SCPI Programming Reference
280
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...