Entering Calibration Values and Storing Calibration Constants
Selecting the Calibration Mode
Use the CALibration:ADC? query to begin ADC calibration. The response to this query indicates a suc-
cessful calibration (0) or a failure (1). Following the ADC calibration, use the measurement commands to
set the DMM measurement parameters and functions for the mode being calibrated.
Front Panel: [Utility] > Test/Admin > Calibrate > Perform Cal > Cal Value > Perform ADC Cal
Entering Calibration Values
To enter a calibration value from the remote interface use CALibration:VALue <
value
> followed by the
CALibration? query. The response to the query indicates whether the calibration is valid.
Front Panel: [Utility] > Test/Admin > Calibrate > Perform Cal > Cal Value > (enter value) > Perform
Cal Step
Storing the Calibration Constants
To store calibration values in non-volatile memory, use CALibration:STORe.
Front Panel: [Utility] > Test/Admin > Calibrate > Save Cal Values
You must have performed a successful calibration immediately before pressing
Save Cal Values
.
Calibration Adjustment Procedures
530
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...