Range, Resolution and Integration Time (shown in Aperture and NPLCs)
The following three tables show the integration time (in PLC and seconds) for each range and resolution.
To determine the PLC for a specified range and resolution:
1.
Locate the range in the left column. Measurement units do not apply (i.e. for 1 V or 1 A, use the 1
range; for 100 V or 100 Ω, use the 100 range).
2.
From the range value, follow the corresponding row of resolutions to the right until the specified res-
olution lies within two adjacent columns for your DMM model.
3.
The NPLC value at the top of the left-adjacent column is the NPLC "setting" for the given range and
resolution.
For example: for the 34461A, a specified 10 VDC range and 10 µV resolution, locate range 10 in the table.
Moving to the right, the 10 µV resolution (1E-5) is found in the column under 10 NPLC.
Model
34460A
34461A
PLC
100
10
1
0.2
0.02
100
10
1
0.2
0.02
Aperture (60 Hz Power) 1.67s 0.167s 16.7ms
3ms
0.3ms
1.67s
0.167s 16.7ms
3ms
0.3ms
Aperture (50 Hz Power)
2s
0.2s
20ms
2s
0.2s
20ms
ResFactor
1
3ppm 10ppm 30ppm 100ppm 300ppm 0.3ppm 1ppm
3ppm 10ppm 100ppm
Range
Resolution
1 m
3E-9
1E-8
3E-8
1E-7
3E-7
3E-10
1E-9
3E-9
1E-8
1E-7
10 m
3E-8
1E-7
3E-7
1E-6
3E-6
3E-9
1E-8
3E-8
1E-7
1E-6
100 m
3E-7
1E-6
3E-6
1E-5
3E-5
3E-8
1E-7
3E-7
1E-6
1E-5
1
3E-6
1E-5
3E-5
1E-4
3E-4
3E-7
1E-6
3E-6
1E-5
1E-4
10
3E-5
1E-4
3E-4
1E-3
3E-3
3E-6
1E-5
3E-5
1E-4
1E-3
100
3E-4
1E-3
3E-3
1E-2
3E-2
3E-5
1E-4
3E-4
1E-3
1E-2
1 k
3E-3
1E-2
3E-2
1E-1
3E-1
3E-4
1E-3
3E-3
1E-2
1E-1
10 k
3E-2
1E-1
3E-1
1
3
3E-3
1E-2
3E-2
1E-1
1
100 k
3E-1
1
3
10
30
3E-2
1E-1
3E-1
1
10
1 M
3
10
30
100
300
3E-1
1
3
10
100
10 M
30
100
300
1 k
3 k
3
10
30
100
1 k
100 M
300
1 k
3 k
10 k
30 k
30
100
300
1 k
10 k
1
ResFactor x Range = Resolution.
SCPI Programming Reference
452
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...