Select the Elog Trigger Source
The TRIGger:ELOG command generates an immediate trigger regardless of the trigger source. Unless
you are using this command, select a trigger source from the following:
Trigger
Source
Description
Bus
Selects GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
External
Selects ANY pin that is configured as a Trigger Input on the digital control port.
Immediate
Triggers the transient as soon as it is INITiated.
Pin<1-7>
Selects a specific pin that is configured as a Trigger Input on the digital control port.
Use the following commands to select one of the available trigger sources:
Front Panel Menu Reference
SCPI Command
Not available
To select Bus triggers:
To select digital pin 5 as the trigger:
To select expression1 as the trigger:
Initiate and Trigger the Elog
When the power supply is turned on, the trigger system is in the Idle state. In this state, the trigger
system is disabled, ignoring all triggers. The INITiate command enables the measurement system to
receive triggers. To initiate and trigger the Elog:
Front Panel Menu Reference
SCPI Command
Not available
To initiate the Elog:
To trigger the Elog:
Alternatively, if the trigger source is BUS, you can also
program a
or an IEEE-488 <get> command.
When triggered, the Elog starts placing data in the internal measurement buffer. Because the buffer is
only large enough to hold 20 seconds of accumulated measurement your PC application must
periodically retrieve (or fetch) the data from this buffer.
4 Using the Advanced Power System
160
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...