![Keysight Technologies N6900 Series Скачать руководство пользователя страница 278](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217278.webp)
Parameter
Typical Return
(none)
<time>
Return the number of triggers: TRIG:ACQ:IND:COUN?
TRIGger:ACQuire:TOUTput[:ENABle] 0|OFF|1|ON
TRIGger:ACQuire:TOUTput[:ENABle]?
Enables measurement triggers to be sent to a digital port pin. The digital port pin must be configured
as trigger output before it can be used as a trigger source (see
).
Parameter
Typical Return
0|OFF|1|ON , *RST OFF
0 or 1
Enable sending measurement triggers to digital pins: TRIG:ACQ:TOUT ON
TRIGger:ACQuire:SOURce <
source
>
TRIGger:ACQuire:SOURce?
Selects the trigger source for the acquisition system:
BUS
Selects a remote interface trigger command.
CURRent1
Selects an output current level.
EXTernal
Selects ALL digital port pins that have been configured as trigger sources.
EXPRession<1-8>
Selects a user-defined expression as the trigger source.
PIN<1-7>
Selects a digital port pin configured as a trigger input.
TRANsient1
Selects the transient system as the trigger source.
VOLTage1
Selects an output voltage level.
Parameter
Typical Return
BUS|CURRent1 | EXTernal | EXPRession<1-8> |
PIN<1-7> | TRANsient1 | VOLTage1, *RST BUS
BUS, CURR1, EXT, EXPR<n>, PIN<n>, TRAN1,
or VOLT1
Select digital port pin 1 as the measurement trigger source:
TRIG:ACQ:SOUR PIN1
TRIGger:ARB:SOURce <
source
>
TRIGger:ARB:SOURce?
Selects the trigger source for arbitrary waveforms:
BUS
Selects a remote interface trigger command.
EXTernal
Selects ALL digital port pins that have been configured as trigger sources.
EXPRession<1-8>
Selects a user-defined expression as the trigger source.
IMMediate
Triggers the transient as soon as it is INITiated.
PIN<1-7>
Selects a digital port pin configured as a trigger input.
Keysight N6900/N7900 Series Operating and Service Guide
278
6 SCPI Programming Reference
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...