![Keysight Technologies N6900 Series Скачать руководство пользователя страница 275](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217275.webp)
SYSTem:SIGNal:DEFine EXPRession<1-8>, <"expression">
SYSTem:SIGNal:DEFine? EXPRession<1-8>
Defines up to eight signal expressions. Refer to
for a complete list of signal
sources and operators that are used to "build" each signal expression.
Parameter
Typical Return
EXPRession<1-8>
(none)
<"expression">
<"expression">
Defines the first expression to be true when the output is on: SYST:SIGN:DEF EXPR1, "OutpOn"
Expressions must be enclosed in quotes ("") inside SCPI commands. Expressions are case-insensitive. You can enter alpha char-
acters as all caps , all lowercase, or mixed case. Spaces must be included on either side of the Boolean operators And, Or , and
Not. Spaces are not required around parentheses.
SYSTem:TEMPerature:AMBient?
Returns the temperature at the air inlet. Values are returned in degrees C.
Parameter
Typical Return
(none)
<"ambient temp">
Return the ambient air temperature: SYST:TEMP:AMB?
SYSTem:TIME <hh>,<mm>,<ss>
SYSTem:TIME?
Sets the time of the system clock. Specify hours (0 to 23), minutes (0 to 59), and seconds (0 to 59).
Parameter
Typical Return
<hh>,<mm>,<ss>
<hh,mm,ss>
Set the clock to 8:30 PM: SYST:TIME 20,30,0
l
Setting the time causes the optional BBR logging function to stop and restart using the new time.
SYSTem:VERSion?
Returns the SCPI version that the instrument complies with. Cannot be determined from front panel.
Parameter
Typical Return
(none)
<"YYYY.V">
Return the SCPI version: SYST:VERS?
l
In the return string, YYYY represents the year and V represents the version.
6 SCPI Programming Reference
275
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...