![Keysight Technologies N6900 Series Скачать руководство пользователя страница 328](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217328.webp)
N6951A/N6971A with Option 301
N6951A/N6971A Test Record
Report Number ______
Date__________
Test Description
Model
Min. Specs
Results
Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
+ 18 mV
Vout – 2 mV
19.992 V
Vout – 8 mV
_______
_______
_______
_______
+ 22 mV
Vout + 2 mV
20.008 V
Vout + 8 mV
CV Load Effect:
Both
– 0.75 mV
_______
+ 0.75 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 9 mV
+ 1 mV
Transient response @ 100 µs:
Both
– 150 mV
_______
+ 150 mV
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
N6951A
N6971A
N6951A
N6971A
N6951A
N6971A
N6951A
N6971A
– 8 mA
– 15 mA
Iout – 8 mA
Iout – 15 mA
49.972 A
99.945 A
Iout – 0.028 A
Iout – 0.055 A
_______
_______
_______
_______
_______
_______
_______
_______
+ 8 mA
+ 15 mA
Iout + 8 mA
Iout + 15 mA
50.028 A
100.055 A
Iout + 0.028 A
Iout + 0.055 A
Low Current Measurement Range:
N6951A
N6971A
Iout – 0.0035 A
Iout – 0.007 A
_______
_______
Iout + 0.0035 A
Iout + 0.007 A
CC Load Effect:
N6951A
N6971A
– 3 mA
– 6 mA
_______
_______
+ 3 mA
+ 6 mA
Current Sink Tests
10% of current rating:
100% of current rating:
N6951A
N6971A
N6951A
N6971A
– 5.01 A
– 10.019 A
– 50.028 A
– 100.055 A
_______
_______
_______
_______
– 4.99 A
– 9.981 A
– 49.972 A
– 99.945 A
N6951ASettings
N6971ASettings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
Low Current Measurement Range:
CC Load Effect:
10% of current rating (Isink):
100% of current rating (Isink):
Voltage priority: 20 mV, 50 A
Voltage priority: 20 V, 50 A
Voltage priority: 20 V, 50 A
Voltage priority: 20 V, 25 A to 50 A
Current priority: 20 V, 0 A
Current priority: 20 V, 50 A
Current priority: 20 V, 5 A
Current priority: 20 V, 50 A
Current priority: 20 V, - 5 A
External supply: 20 V, 5.5 A
Current priority: 20 V, - 50 A
External supply: 20 V, 55 A
Voltage priority: 20 mV, 100 A
Voltage priority: 20 V, 100 A
Voltage priority: 20 V, 100 A
Voltage priority: 20 V, 50 A to 100 A
Current priority: 20 V, 0 A
Current priority: 20 V, 100 A
Current priority: 20 V, 10 A
Current priority: 20 V, 100 A
Current priority: 20 V, - 10 A
External supply: 20 V, 11 A
Current priority: 20 V, - 100 A
External supply: 20 V, 110 A
7 Verification and Calibration
328
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...