FETCh[:SCALar]:CURRent:MAXimum?
FETCh[:SCALar]:VOLTage:MAXimum?
FETCh[:SCALar]:POWer:MAXimum?
FETCh[:SCALar]:CURRent:MINimum?
FETCh[:SCALar]:VOLTage:MINimum?
FETCh[:SCALar]:POWer:MINimum?
Returns the maximum or minimum value. Values returned are either in amperes, volts, or watts.
Parameter
Typical Return
(none)
<MIN value> or <MAX value>
Returns the measured maximum current FETC:CURR:MAX?
Returns the measured maximum voltage FETC:VOLT:MAX?
Returns the measured minimum power FETC:POW:MIN?
l
FETC:POW:MIN? and FETC:POW:MAX? are only available on units with firmware revisions
A.02.01.1256 and B.02.01.1256 and up. See
FETCh:AHOur? [IGNORE_OVLD]
FETCh:WHOur? [IGNORE_OVLD]
FETCh:AHOur? Returns the accumulated amp-hours.
FETCh:WHOur? - Returns the accumulated watt-hours.
Refer to
Amp-Hour and Watt-Hour Measurements
for details.
If any measurement sample was overrange, the query returns SCPI Not a Number (9.91E37). If the
optional IGNORE_OVLD parameter is sent, the accumulated measurement will be returned even if
some samples were outside of the measurement range.
Parameter
Typical Return
IGNORE_OVLD ignore overload measurements
<amp-hours> or <watt-hours>
Returns the amp-hour measurement FETC:AHO?
FETCh:ARRay:CURRent[:DC]? [<
start_index
>, <
points
>]
FETCh:ARRay:VOLTage[:DC]? [<
start_index
>, <
points
>]
FETCh:ARRay:POWer[:DC]? [<
start_index
>, <
points
>]
Returns the instantaneous measurement. Values returned are either in amperes, volts, or watts.
Optional parameters specify a subset starting at <startindex> and of length <points>.
The return format depends on the settings of the FORMat:BORDer and FORMat[:DATA] commands.
When the data format is set to ASCII, returned values are comma separated. When the data format is
set to REAL, data is returned as single precision floating point values in definite length arbitrary block
response format.
Keysight N6900/N7900 Series Operating and Service Guide
228
6 SCPI Programming Reference
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...