To offset the beginning of the acquisition buffer relative to the acquisition trigger:
Front Panel Menu Reference
SCPI Command
Select Measure\Sweep.
Enter an Offset value. Then press Select.
To offset the measurement by 100 points use:
When the value is 0, all measurement samples are taken after the trigger. Positive values represent
the delay after the trigger occurs but before the samples are acquired. This can be used to exclude
measurement samples that occur during the delay time. (Delay time = offset x sample period).
Negative values represent data samples taken prior to the trigger. This lets you acquire measurement
samples prior to the trigger.
Select the trigger source
A TRIGger:ACQuire[:IMMediate] command over the bus will always generate an
immediate measurement trigger, regardless of the selected trigger source.
Unless you are using TRIGger:ACQuire[:IMMediate], select a trigger source from the following:
Trigger Source
Description
Bus
Selects GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
Current1
Selects an output current level.
Expression<1-8>
Selects one of eight user-defined expressions. See
Using Expression Signal Routing
External
Selects ANY pin that is configured as a Trigger Input on the digital control port.
Pin<1-7>
Selects a specific pin that is configured as a Trigger Input on the digital control port.
Transient1
Selects the unit’s transient system. You must also set up the transient system to gen-
erate a trigger out signal. See
Voltage1
Selects an output voltage level.
4 Using the Advanced Power System
138
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...