Using Expression Signal Routing
This section describes how expressions can be used to program signal routing. You can also use the
to program signal routing. For an overview of how the expression signals are
mapped into the trigger system, refer to
Configuring Threshold Comparators
Specifying Signal Expression Targets
Signal Routing Overview
You can configure up to eight flexible, general purpose signal routing expressions. These signal
expressions can be used to drive digital port pins, trigger measurements and transients, change
output state, generate user-defined protection and user-defined status signals. Expressions are built
using a large selection of signal inputs along with Boolean operators and programmable delays.
The following figure illustrates the signal routing paths:
4 Using the Advanced Power System
142
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...