In this example, pin 6 is configured as the output On control. Pin 7 is configured as the output Off
control. The ground or Common pins are connected together.
Front Panel Menu Reference
SCPI Command
Select System\IO\DigPort\Pins.
Select pin 6, then Function, then On Couple.
Select Pins., select pin7, then Function, then Off Couple.
Repeat these steps for units #2 and #3
To set pin 6 of unit 1 as the ON control:
To configure pin 7 of unit 1 as the OFF control:
Repeat these commands for units 2 and 3.
Once configured and enabled, turning the output on or off on any coupled unit will cause all coupled
units to turn on or off according to their user-programmed delays.
Keysight N6900/N7900 Series Operating and Service Guide
157
4 Using the Advanced Power System
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...