![Keysight Technologies N6900 Series Скачать руководство пользователя страница 325](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217325.webp)
N6953A/N6973A
N6953A/N6973A Test Record
Report Number ______
Date__________
Test Description
Model
Min. Specs
Results
Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
+ 51 mV
Vout – 9 mV
59.973 V
Vout – 27 mV
_______
_______
_______
_______
+ 69 mV
Vout + 9 mV
60.027 V
Vout + 27 mV
CV Load Effect:
Both
– 2 mV
_______
+ 2 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 9 mV
+ 1 mV
Transient response @ 100 µs:
Both
– 150 mV
_______
+ 150 mV
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
N6953A
N6973A
N6953A
N6973A
N6953A
N6973A
N6953A
N6973A
– 5 mA
– 10 mA
Iout – 5 mA
Iout – 10 mA
16.6783 A
33.2567 A
Iout – 0.0217 A
Iout – 0.0433 A
_______
_______
_______
_______
_______
_______
_______
_______
+ 5 mA
+ 10 mA
Iout + 5 mA
Iout + 10 mA
16.7217 A
33.3433 A
Iout + 0.0217 A
Iout + 0.0433 A
CC Load Effect:
N6953A
N6973A
– 1 mA
– 1.5 mA
_______
_______
+ 1 mA
+ 1.5 mA
Current Sink Tests
10% of current rating:
100% of current rating:
N6953A
N6973A
N6953A
N6973A
– 1.6767 A
– 3.3433 A
– 16.7217 A
– 33.3433 A
_______
_______
_______
_______
– 1.6633 A
– 3.3167 A
– 16.6783 A
– 33.2567 A
N6953A Settings
N6973A Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
10% of current rating (Isink):
100% of current rating (Isink):
Voltage priority: 60 mV, 16.7 A
Voltage priority: 60 V, 16.7 A
Voltage priority: 60 V, 16.7 A
Voltage priority: 60 V, 8.35 A to 16.7 A
Current priority: 60 V, 0 A
Current priority: 60 V, 16.7 A
Current priority: 60 V, 16.7 A
Current priority: 60 V, - 1.67 A
External supply: 60 V, 1.837 A
Current priority: 60 V, - 16.7 A
External supply: 60 V, 18.37 A
Voltage priority: 60 mV, 33.3 A
Voltage priority: 60 V, 33.3 A
Voltage priority: 60 V, 33.3 A
Voltage priority: 60 V, 16.7 A to 33.3 A
Current priority: 60 V, 0 A
Current priority: 60 V, 33.3 A
Current priority: 60 V, 33.3 A
Current priority: 60 V, - 3.33 A
External supply: 60 V, 3.66 A
Current priority: 60 V, - 33.3 A
External supply: 60 V, 36.6 A
Keysight N6900/N7900 Series Operating and Service Guide
325
7 Verification and Calibration
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...