Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”. Enter the data from
the external DMM. This should be about 70% of the full-scale
current rating. Press Enter when done. Press Back to finish.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Current Low Range Measurement
Step 1.Disconnect all equipment form the output terminals.
Step 2. Select the current low range measurement calibration.
Front Panel Menu Reference
SCPI Command
Select System\Admin\Cal\Curr\Imeas.
Check that the shunt is connected and select Next.
Specify the low range measurement. Low ranges vary
by model. This selects the 2 A range:
Step 3. Select the first current calibration point. Wait 5 minutes for the temperature to settle.
Front Panel Menu Reference
SCPI Command
Wait 5 minutes, then select Next again.
Step 4. Connect a precision shunt resistor to the output. The shunt resistor should be able to measure
the full-scale current of the low range (see
). Connect the Keysight 3458A DMM across the
shunt resistor.
Front Panel Menu Reference
SCPI Command
Check that the shunt is connected and select
Next.
Not applicable
Step 5. Select the second current calibration point. There is no wait required for the second point.
Calculate the shunt current (I=V/R) and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”. Enter the data from
the external DMM. This should be about 100% of the full-
scale low range rating. Press Enter when done. Press Back to
finish.
CAL:LEV P2
*OPC? CAL:DATA <data>
Current Sharing Calibration
This procedure calibrates the Imon signal that is used when units are connected in parallel.
Step 1. Connect a short across the + and - output terminals. Connect the Keysight 3458A DMM across
pins 2 and 3 of the Sharing connector (see
).
Step 2. Select the current sharing calibration.
7 Verification and Calibration
320
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...