![Keysight Technologies N6900 Series Скачать руководство пользователя страница 216](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217216.webp)
Parameter
Typical Return
<
"date"
> String program data.
<last cal date>
Enters the calibration date: CAL:DATE "12/12/12"
CALibrate:LEVel P1|P2|P3
CALibrate:LEVel?
Advances to the next level in the calibration. P1 is the first level; P2 is the second; P3 is the third.
Parameter
Typical Return
P1|P2|P3
(none)
Selects the first calibration point: CAL:LEV P1
l
Some calibration sequences may require some settling time after sending CAL:LEV but before
reading the data from the DVM and sending CAL:DATA.
CALibrate:PASSword <
password
>
Sets a numeric password to prevent unauthorized calibration. This is the same as the
password.
Parameter
Typical Return
<
password
>
A numeric value of up to 15 digits
(none)
Set a new password to a value of 1234: CAL:PASS 1234
l
If the password is set to 0, password protection is removed and the ability to enter calibration
mode is unrestricted. The as-shipped setting is 0 (zero).
l
To change the password: unsecure calibration memory with old code, then set the new code.
l
If you forget your password, refer to
l
This setting is non-volatile; it will not be changed by power cycling or *RST.
CALibrate:RESistance:BOUT
Calibrates the bottom out resistance.
Parameter
Typical Return
(none)
(none)
Calibrates the bottom out resistance: CAL:RES:BOUT
Keysight N6900/N7900 Series Operating and Service Guide
216
6 SCPI Programming Reference
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...