![Keysight Technologies N6900 Series Скачать руководство пользователя страница 340](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217340.webp)
N7977A and N6977A with Option 301
N7977A/N6977A Option 301 Test Record
Report Number ______
Date__________
Test Description
Model
Min. Specs
Results
Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
+ 146 mV
Vout – 14 mV
159.938 V
Vout – 62 mV
_______
_______
_______
_______
+ 174 mV
Vout + 14 mV
160.062 V
Vout + 62 mV
CV Load Effect:
Both
– 4 mV
_______
+ 4 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 9 mV
+ 1 mV
Transient response @ 100 µs:
Both
– 400 mV
_______
+ 400 mV
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
– 2 mA
Iout – 2 mA
12.493 A
Iout – 0.007 A
_______
_______
_______
_______
+ 2 mA
Iout + 2 mA
12.507 A
Iout + 0.007 A
Low Current Measurement Range:
Both
Iout – 0.875 mA
_______
Iout + 0.875 mA
CC Load Effect:
Both
– 0.8 mA
_______
+ 0.8 mA
Current Sink Tests
10% of current rating:
100% of current rating:
Both
Both
– 1.2525 A
– 12.507 A
_______
_______
– 1.2475 A
– 12.493 A
N6977A/N6977A Option 301 Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
Low Current Measurement Range:
CC Load Effect:
10% of current rating (Isink):
100% of current rating (Isink):
Voltage priority: 160 mV, 12.5 A
Voltage priority: 160 V, 12.5 A
Voltage priority: 160 V, 12.5 A
Voltage priority: 160 V, 6.25 A to 12.5 A
Current priority: 160 V, 0 A
Current priority: 160 V, 12.5 A
Current priority: 160 V, 1.25 A
Current priority: 160 V, 12.5 A
Current priority: 160 V, - 1.25 A
External supply: 160 V, 1.375 A
Current priority: 160 V, - 12.5 A
External supply: 160 V, 13.75 A
7 Verification and Calibration
340
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...