![Keysight Technologies N6900 Series Скачать руководство пользователя страница 225](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217225.webp)
DISPlay Subsystem
Display commands control the front panel display.
DISPlay[:WINDow][:STATe] 0|OFF|1|ON
DISPlay[:WINDow][:STATe]?
Turns the front panel display on or off.
Parameter
Typical Return
0|OFF|1|ON, *RST ON
0 or 1
Turns the front panel display off: DISP OFF
DISPlay[:WINDow]:VIEW METER_VI|METER_VP|METER_VIP
DISPlay[:WINDow]:VIEW?
Selects the parameters to display on the front panel.
METER_VI displays output voltage and current.
METER_VP displays output voltage and power.
METER_VIP displays output voltage, current, and power.
Parameter
Typical Return
METER_VI|METER_VP|METER_VIP, *RST METER_VI
METER_VI, METER_VP, or METER_VIP
To display voltage and power: DISP:VIEW METER_VP
DISPlay:SAVer[:STATe] 0|OFF|1|ON
DISPlay:SAVer[:STATe]?
Turns the front panel screen saver on or off.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Turns the front panel screen saver on: DISP:SAV ON
6 SCPI Programming Reference
225
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...