When a bit in the PTR register is set to 1, then a 0-to-1 transition of the corresponding bit in the
Operation Condition register causes that bit in the Operation Event register to be set.
STATus:PRESet sets all bits in the PTR registers and clears all bits in the NTR registers.
Parameter
Typical Return
A decimal value corresponding to the binary-weighted sum of
the register's bits.
<bit value>
Enable bit 3 and 4 in the NTR register: STAT:OPER:NTR 24
Enable bit 3 and 4 in the PTR register: STAT:OPER:PTR 24
l
If the same bits in both NTR and PTR registers are set to 1, then any transition of that bit at the
Operation Condition register sets the corresponding bit in the Operation Event register.
l
If the same bits in both NTR and PTR registers are set to 0, then no transition of that bit at the Oper-
ation Condition register can set the corresponding bit in the Operation Event register.
l
The value returned is the binary-weighted sum of all bits set in the register.
STATus:OPERation:USER<1|2>:SOURce EXPRession<1-8>|NONE
STATus:OPERation:USER<1|2>:SOURce?
Maps the specified expression to one of the two user-defined
bits (USER1 or USER2).
When the specified expression is true, the selected user status bit will be set.
Parameter
Typical Return
EXPRession<1-8>|NONE *RST NONE
EXPR<1-8>
Maps a user-defined status bit to an expression:
STAT:OPER:USER1:SOUR EXPR1
STATus:PRESet
Presets all Enable, PTR, and NTR registers.
Operation register
Questionable register
Preset setting
STAT:OPER:ENAB
STAT:QUES<1|2>:ENAB
all defined bits are disabled
STAT:OPER:NTR
STAT:QUES<1|2>:NTR
all defined bits are disabled
STAT:OPER:PTR
STAT:QUES<1|2>:PTR
all defined bits are enabled
Parameter
Typical Return
(none)
(none)
Preset the Operation and Questionable registers: STAT:PRES
Keysight N6900/N7900 Series Operating and Service Guide
266
6 SCPI Programming Reference
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...