![Keysight Technologies N6900 Series Скачать руководство пользователя страница 239](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217239.webp)
INITiate Subsystem
Initiate commands initialize the trigger system. This moves the trigger system from the "idle" state to
the "wait-for- trigger" state; which enables the instrument to receive triggers. An event on the
selected trigger source causes the trigger to occur.
INITiate[:IMMediate]:ACQuire
INITiate[:IMMediate]:ELOG
INITiate[:IMMediate]:TRANsient
INITiate:ACQuire - Initiates the measurement trigger system.
INITiate:ELOG - Initiates external data logging.
INITiate:TRANsient - Initiates the transient trigger system.
Parameter
Typical Return
(none)
(none)
Initiate the measurement-trigger system: INIT:ACQ
l
It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving
the INITiate command.
l
If a trigger occurs before the trigger system is ready for it, the trigger will be ignored. Check the
WTG_meas bit in the operation status register to know when the instrument is ready.
l
Use ABORt commands to return the instrument to Idle.
INITiate:CONTinuous:TRANsient 0|OFF|1|ON
INITiate:CONTinuous:TRANsient?
Continuously initiates the transient trigger system. This allows multiple triggers to generate multiple
output transients.
Parameter
Typical Return
0|OFF|1|ON, *RST ON
0 or 1
Continually initiates the output trigger system: INIT:CONT:TRAN ON
l
With initiate continuous disabled, the output trigger system must be initiated for each trigger using
the INITiate:TRANsient command.
l
ABORt:TRANsient does not turn off continuous triggers if INITiate:CONTinuous:TRANsient ON has
been programmed. In this case, the trigger system will automatically re-initiate.
6 SCPI Programming Reference
239
Keysight N6900/N7900 Series Operating and Service Guide
Содержание N6900 Series
Страница 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Страница 2: ......
Страница 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Страница 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 90: ...Rack Mounting 2 Installing the Instrument 90 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 289: ...Status Diagram 6 SCPI Programming Reference 289 Keysight N6900 N7900 Series Operating and Service Guide...
Страница 354: ...Cover screws 8 Service and Maintenance 354 Keysight N6900 N7900 Series Operating and Service Guide...