N5511A Phase Noise Test System User’s Guide 99
Absolute Measurement Examples
Input Ports
Input Ports
The N5511A Phase Noise Tests System phase detector frequency specification
are the following.
For optimal performance, however, when performing a measurement, N5510
software defaults to detector selections that do not match the hardware
capabilities. When performing a measurement, the software default detector
selections use the following guidelines:
This can be overwritten by manually selecting the detector to be utilized in the
measurement setup, as long as the input signal is within the hardware
specifications of the detector input.
Low Frequency Inputs
(SMA)
All Options
High Frequency Inputs
(Type K)
Option 526
High Frequency Inputs
(Type K)
Option 540
AM Noise Input
(SMA)
All Options
Carrier frequency
range
50 kHz to 3 GHz
1.2 GHz to 26.5 GHz
1.2 GHz to 40 GHz
50 kHz to 40 GHz
RF input power
LO input power
0 dBm to +23 dBm
+15 dBm to +23 dBm
0 dBm to +15 dBm
+7 dBm to +15 dBm
0 dBm to +15 dBm
+7 dBm to +15 dBm
0 dBm to +30 dBm
N/A
Low Frequency Inputs
50 kHz to 1.6 GHz
High Frequency Inputs
1.6 GHz and above
Содержание N5511A
Страница 1: ...User s Guide Keysight N5511A Phase Noise Test System...
Страница 4: ...4...
Страница 11: ...Contents N5511A Phase Noise Test System User s Guide 11...
Страница 12: ...12 N5511A Phase Noise Test System User s Guide Contents...
Страница 38: ...38 N5511A Phase Noise Test System User s Guide Phase Noise Basics What is Phase Noise...
Страница 54: ...54 N5511A Phase Noise Test System User s Guide Expanding Your Measurement Experience GUI Features...
Страница 166: ...166 N5511A Phase Noise Test System User s Guide Absolute Measurement Examples OCXO Dual Channel Measurement...