![Keysight Technologies N5511A Скачать руководство пользователя страница 268](http://html1.mh-extra.com/html/keysight-technologies/n5511a/n5511a_user-manual_1962254268.webp)
268 N5511A Phase Noise Test System User’s Guide
AM Noise Measurement Fundamentals
Calibration and Measurement General Guidelines
Figure 11-12
Dual Channel AM noise measurement setup using single-sided spur
2.
Measure the power which will be applied to the AM detector. It must be
between 0 and +23 dBm.
3.
Measure the carrier-to-single-sided-spur ratio and the spur offset at the
input to the AM detector with an RF spectrum analyzer. See
.
The spur should be adjusted such that it is between –30 and –60 dBc, with
a carrier offset of 10 Hz to 20 MHz.
Figure 11-13
Measuring relative spur level
4.
Enter the relative power level and offset of the tone into the measurement
definition in the software. The AM detector will remove the PM
component, thus the end result in the measurement will be a spur that is 6
dB lower than what the analyzer measured. See
Содержание N5511A
Страница 1: ...User s Guide Keysight N5511A Phase Noise Test System...
Страница 4: ...4...
Страница 11: ...Contents N5511A Phase Noise Test System User s Guide 11...
Страница 12: ...12 N5511A Phase Noise Test System User s Guide Contents...
Страница 38: ...38 N5511A Phase Noise Test System User s Guide Phase Noise Basics What is Phase Noise...
Страница 54: ...54 N5511A Phase Noise Test System User s Guide Expanding Your Measurement Experience GUI Features...
Страница 166: ...166 N5511A Phase Noise Test System User s Guide Absolute Measurement Examples OCXO Dual Channel Measurement...