N5511A Phase Noise Test System User’s Guide 81
Absolute Measurement Fundamentals
Inserting a Device
The N5511A's dual channel cross-correlation capability offers a unique
advantage when performing measurements requiring amplifiers. In a dual
channel configuration, amplifiers of identical gain can be placed in each of the
separate channel paths. See
.
Figure 5-19
Dual Channel Configuration
When a measurement is configured in such manner, the noise contribution
from the amplifiers is removed by cross-correlation due to their noise being
uncorrelated. This enables the resulting noise floor measured to be the true
performance of the DUT. This technique also means that the performance of
the amplifiers need not be a critical factor; however, one must be considerate
of the impact that higher noise amplifiers will have on the number of
cross-correlations required to reach the correlated noise floor of the DUT.