266 N5511A Phase Noise Test System User’s Guide
AM Noise Measurement Fundamentals
Calibration and Measurement General Guidelines
Figure 11-10
AM Detector removes carrier
4.
Enter the carrier-to-sideband ratio and offset frequency, then measure the
calibration constant.
5.
Remove the AM source and reconnect the DUT.
6.
Measure noise data and interpret the results.
The quadrature meter should be at zero volts due to the blocking capacitor at the AM detector’s
output.