![Keysight Technologies N5511A Скачать руководство пользователя страница 315](http://html1.mh-extra.com/html/keysight-technologies/n5511a/n5511a_user-manual_1962254315.webp)
N5511A Phase Noise Test System User’s Guide 315
Advanced Software Features
Phase-Lock-Loop Suppression
There are four different curves available for this graph:
a.
“Measured” loop suppression curve (
)—this is the result of the
loop suppression measurement performed by the N5511A system.
b.
“Smoothed” measured suppression curve (
)—this is a curve-fit
representation of the measured results, it is used to compare with the
“theoretical” loop suppression.
c.
“Theoretical” suppression curve (
)—this is the predicted loop
suppression based on the initial loop parameters defined/selected for this
particular measurement (kphi, kvco, loop bandwidth, filters, gain, etc.).
d.
“Adjusted” theoretical suppression curve (
through
) —this is the new “adjusted” theoretical value of suppression for this
measurement. It is based on changing loop parameters (in the theoretical
response) to match the “smoothed” measured curve as closely as possible.
Содержание N5511A
Страница 1: ...User s Guide Keysight N5511A Phase Noise Test System...
Страница 4: ...4...
Страница 11: ...Contents N5511A Phase Noise Test System User s Guide 11...
Страница 12: ...12 N5511A Phase Noise Test System User s Guide Contents...
Страница 38: ...38 N5511A Phase Noise Test System User s Guide Phase Noise Basics What is Phase Noise...
Страница 54: ...54 N5511A Phase Noise Test System User s Guide Expanding Your Measurement Experience GUI Features...
Страница 166: ...166 N5511A Phase Noise Test System User s Guide Absolute Measurement Examples OCXO Dual Channel Measurement...