N5511A Phase Noise Test System User’s Guide 181
Residual and Additive Noise Measurement Fundamentals
Calibrating the Measurement
The DC peak voltage method is a very convenient calibration for residual
measurements as a cal source is not needed. The user adjusts the phase shift
until a peak positive voltage is measured and then adjusted again to measure a
minimum negative voltage. The meters on the application are used to identify
these peaks.
shows how the meters operate as well as a
mathematical representation of ± DC peak voltage cal. Measuring the peak
positive voltage means adjusting the phase shifter(s) so that the LO path of
each channel is in phase with the RF path into the mixer. To measure the
minimum negative voltage, the phase shifter(s) are adjusted so that the LO
path into the phase detector is 180 degrees out of phase relative to the RF
input.
Figure 7-11
± DC Peak Voltage
This method has moderate accuracy since it is measuring the signal in the IF
after the mixer, at which point the signal present may not be a perfect sinusoid.
This causes the peak value measured to be inaccurate which can affect the
Kphi by as much as 2 dB. In many cases this may not be an issue.
Содержание N5511A
Страница 1: ...User s Guide Keysight N5511A Phase Noise Test System...
Страница 4: ...4...
Страница 11: ...Contents N5511A Phase Noise Test System User s Guide 11...
Страница 12: ...12 N5511A Phase Noise Test System User s Guide Contents...
Страница 38: ...38 N5511A Phase Noise Test System User s Guide Phase Noise Basics What is Phase Noise...
Страница 54: ...54 N5511A Phase Noise Test System User s Guide Expanding Your Measurement Experience GUI Features...
Страница 166: ...166 N5511A Phase Noise Test System User s Guide Absolute Measurement Examples OCXO Dual Channel Measurement...