N5511A Phase Noise Test System User’s Guide 189
Residual Measurement Examples
Amplifier Measurement Example - Single Channel
Configuring Equipment
Set Up stimulus PSG
Frequency: 1 GHz
Amplitude: 14 dBm
Set SSB Cal Tone
Set the cal source to be at a 1 kHz offset from the stimulus carrier at the output of the coupler
and
−
40 dB relative to the carrier. A signal analyzer should be used to verify the cal tone.