N5511A Phase Noise Test System User’s Guide 73
Absolute Measurement Fundamentals
Tracking Frequency Drift
Evaluating beat note drift
The Checking the beat note section included in each phase lock loop
measurement example in this chapter provides a procedure for adjusting the
beat note to within the Capture Range set for the measurement. If you have not
done so already, verify that the beat note signal can be tuned to within the
Capture Range and that it will remain within the range.
Continue to observe the beat note and verify that it will not drift beyond the
drift tracking range (24% of the PTR) during the measurement period. The
length of the measurement period is primarily a function of the frequency offset
range specified for the measurement (Start to Stop Frequency).
Action
If beat note drift exceeds the limits of the Capture or drift tracking ranges set
for your measurement, the system is not able to complete the measurement.
You have two possible alternatives.
1.
Minimize beat note drift.
—
By Allowing sources to warm-up sufficiently.
—
By Selecting a different reference source with less drift.
2.
Increase the capture and drift tracking Ranges.
—
By Selecting a measurement example in this chapter that specifies a
drift rate compatible with the beat note drift rate you have observed.
—
By Increasing the peak tuning range for the measurement. (Further
information about increasing the PTR is provided in Changing the
PTR.)
Содержание N5511A
Страница 1: ...User s Guide Keysight N5511A Phase Noise Test System...
Страница 4: ...4...
Страница 11: ...Contents N5511A Phase Noise Test System User s Guide 11...
Страница 12: ...12 N5511A Phase Noise Test System User s Guide Contents...
Страница 38: ...38 N5511A Phase Noise Test System User s Guide Phase Noise Basics What is Phase Noise...
Страница 54: ...54 N5511A Phase Noise Test System User s Guide Expanding Your Measurement Experience GUI Features...
Страница 166: ...166 N5511A Phase Noise Test System User s Guide Absolute Measurement Examples OCXO Dual Channel Measurement...