N5511A Phase Noise Test System User’s Guide 115
Absolute Measurement Examples
Single Channel Measurement
peak tuning range for your VCO can be estimated by multiplying the VCO
tuning constant by the tune range of VCO. Refer to
if you are not familiar with the relationship
between the PLL capture range and the peak tuning range of the VCO. The
beat note frequency is set by the relative frequency difference between the two
sources. If you have two very accurate sources set at the same frequency, the
resulting beat note will be very close to 0 Hz. Searching for the beat note will
require that you adjust the center frequency of one of the sources above and
below the frequency of the other source until the beat note appears on the
oscilloscope's display. If incrementing the frequency of one of the sources does
not produce a beat note, you will need to verify the presence of an output
signal from each source before proceeding.
Figure 6-9
Pause point window for when a beat note needs to be tuned
Figure 6-10
Monitor output displaying adjusted beat note within specified range
If the center frequencies of the sources are not close enough to create a beat note within the
capture range, the system will not be able to complete its measurement.
Содержание N5511A
Страница 1: ...User s Guide Keysight N5511A Phase Noise Test System...
Страница 4: ...4...
Страница 11: ...Contents N5511A Phase Noise Test System User s Guide 11...
Страница 12: ...12 N5511A Phase Noise Test System User s Guide Contents...
Страница 38: ...38 N5511A Phase Noise Test System User s Guide Phase Noise Basics What is Phase Noise...
Страница 54: ...54 N5511A Phase Noise Test System User s Guide Expanding Your Measurement Experience GUI Features...
Страница 166: ...166 N5511A Phase Noise Test System User s Guide Absolute Measurement Examples OCXO Dual Channel Measurement...