64
Using the Multi-Wavelength Meter
Calibrating Measurements
3
Press the
CAL
softkey.
4
Press
ELEV
.
5
Use the
and
softkeys to enter the elevation in meters. Entries jump in 500
meter steps from 0 m to 5000 m.
In order for the Agilent 86120B to meet its published specifications, the elevation
value selected with the softkeys must be within 250 meters of the actual elevation.
6
Press
RETURN
to complete the entry.
To select the medium for light
1
Press the Setup key.
2
Press the
MORE
softkey.
3
Press the
CAL
softkey, and make the following selection:
• Press
VACUUM
for wavelengths in a vacuum.
• Press
STD AIR
for wavelengths in standard air.
4
Press
RETURN
to complete the entry.
Converting feet to meters
If you know your elevation in feet, you can convert this value to meters by
using the following equation:
m
ft
3.281
-------------
=
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......