46
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
Measuring broadband devices and chirped lasers
When first turned on (or the green Preset key is pressed), the Agilent 86120B is con-
figured to measure narrowband devices such as DFB lasers and modes of FP lasers.
If you plan to measure broadband devices such as LEDs, optical filters, and chirped
lasers, use the Setup menu first to reconfigure the instrument. When broadband
devices are selected, the display shows the BROAD annotation on the screen.
The measurement algorithm for broadband devices determines the wavelength
based upon the center-of-mass of the power spectrum. The
peak excursion function
is used to determine the value of the integration limits. Care must be taken to ensure
that the integration limits are above any noise. This is especially true when measur-
ing devices with sloping noise floors, like an EDFA amplifier. For more information
on peak excursion,
refer to “Defining Laser-Line Peaks” on page 52
Instrument specifications apply when the Agilent 86120B is configured to measure
narrowband devices. Specifications do not apply when the instrument is configured
to measure broadband devices.
This feature applies to Agilent 86120B instruments with firmware version number
2.0. When first turned on, the instrument briefly displays the firmware version.
Instruments with a firmware version number less than 2.0 do not have this fea-
ture.
To measure broadband devices
1
Press the Setup key.
2
Press
MORE
twice, and then the
DEVICE
softkey.
3
Press the
BROAD
softkey.
To return to measuring narrowband devices, press
NARROW
.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......