Instrument Preset Conditions
Table 8-23. Instrument Preset Conditions (1 of 2)
Item
Settings after Preset
Key Pressed
Settings after Power
Turned On
Display mode
single wavelength
last state
a
Wavelength range limiting
on
last state
Start wavelength
1200 nm
last state
Stop wavelength
1650 nm
last state
Graphical display
off
off
Measurement acquisition
continuous
last state
Wavelength calibration
vacuum
last state
Elevation correction value
not affected
last state
Wavelength units
nm
last state
Amplitude units
dBm
last state
Power offset
0 dB
last state
Peak threshold
10 dB
last state
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......