233
Specifications and Regulatory Information
Definition of Terms
Definition of Terms
Wavelength
Range
refers to the allowable wavelength range of the optical input signal.
Absolute accuracy
indicates the maximum wavelength error over the allowed envi-
ronmental conditions. The wavelength accuracy is based on fundamental physical
constants, which are absolute standards not requiring traceability to artifacts kept at
national standards laboratories. Four He-Ne gas lasers are used. First, there is an
internal 632.991 nm-vacuum (473.6127 THz) reference laser
1
. To verify absolute
wavelength accuracy, three external lasers are measured during manufacturing with
these wavelengths:
730.685 nm, or 410.2896 THz
2
1152.591 nm, or 260.1032 THz
3
1523.488 nm, or 196.7804 THz
3
Differential Accuracy
indicates the maximum wavelength error in measuring the
wavelength difference between two signals that are simultaneously present.
Minimum Resolvable Separation
indicates the minimum wavelength separation of
two laser lines required to measure each wavelength simultaneously. Two laser lines
closer in wavelength than the minimum resolvable separation are not resolved and
one average wavelength is displayed.
Display Resolution
indicates the minimum incremental change in displayed wave-
length.
Amplitude
Calibration Accuracy
indicates the maximum power calibration error at the speci-
fied wavelengths over the allowed environmental conditions. The amplitude calibra-
tion accuracy is traceable to a National Institute of Standards and Technology
(NIST) calibrated optical power meter. NIST is the national standards laboratory of
the United States.
1. Obarski, G. E. 1990. “Wavelength Measurement System for Optical Fiber Communications”
NIST
Technical Note 1336
(February): 18. Take the average of the two frequencies straddling gain center.
2. Moore, C. E. 1971. “Atomic Energy Levels as Derived from the Analysis of Optical Spectra: Vol.
1”
NSRDS-NBS 35, Vol. 1 (COM-72-51282)
(December): 77.
3. D. A. Jennings, F. R. Peterson, and K. M. Evenson. 1979. “Frequency measurement of the 260-
THz (1.15 micron) He-Ne laser”
Optics Letters Vol. 4, No. 5
(May): 129-130.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......