91
Programming
Making Measurements
FETCh does not reconfigure the display. For example, if the display is in the Peak
WL mode, sending :FETCh:ARRay does not configure the display to the List by
WL even though an array of data is returned to the computer.
A common programming error occurs when the :FETCh command is used after an
*RST command. This generates error number –230,
“Data corrupt or stale”
. In this
instance, you must send :INIT:IMM after the *RST command and before :FETCh
command to capture a new array of measurement data.
:CONFigure command
The CONFigure command changes measurement settings without taking a measure-
ment. The instrument is placed in the List by WL, List by Ampl, Peak WL display,
or in the coherence length application.
CONFigure can be queried. The query returns the last configuration setup by the
CONFigure command. The instrument returns a string which is the last instrument
function sent by a CONFigure command or MEASure query. The returned string is
in the short command form. Use caution when using this query, because if any
instrument settings were changed since the last CONFigure command or MEASure
query these changes may not be included in the returned string.
For example, if the last CONFigure command was
:CONFigure:SCALar:POWer:WAVelength 1300NM, MAX
a CONFigure? query would return a string that is similar to the following line:
“POW:WAV 1.300000e-6,0.01”
The
1300NM
and
resolution
values track the actual instrument settings and input
signals. Notice that the quotation marks are part of the returned string.
Return single or multiple measurement values
You can specify whether FETCh, READ, or MEASure returns a single value (SCA-
Lar) or multiple values (ARRay). The following example specifies SCALar data
which returns a single value.
:MEASure:SCALar:POWer:WAVelength? MAX
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......