Using the Multi-Wavelength Meter
Measuring Laser Separation
Channel separation
Suppose that you want to measure separation on a system having the spectrum
shown in the following figure.
The Agilent 86120B displays separation on this spectrum as shown in the following
figure. Notice that the 1541.747 nm laser line is selected as the reference. It is shown
in absolute units. The wavelengths and powers of the remaining responses are
shown relative to this reference. For example, the first response is 2.596 nm below
the reference.
To determine channel spacing, simply read the relative wavelength measurement of
the laser lines immediately preceding and following the reference. Use the
,
,
and
SELECT
softkeys to change the reference laser line and read the channel spac-
ing between each channel.
To measure channel separation
1
Press the front-panel Preset key.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......