Queries the corrected peak data of the input laser line.
Syntax
:CALCulate2:DATA? {FREQuency | POWer | WAVelength | WNUMber}
Constant
Description
FREQuency
Queries the array of laser-line frequencies after the peak search
is completed. If :CALC2:PWAV:STAT is on, the power-
weighted average frequency is returned.
POWer
Queries the array of laser-line powers after the peak search is
completed. If :CALC2:PWAV:STAT is on, the total input power
is returned.
WAVelength
Queries the array of laser-line wavelengths after the peak search
is completed. If :CALC2:PWAV:STAT is on, the power-
weighted average wavelength is returned.
WNUMber
Queries the array of laser-line wave numbers after the peak
search is completed. If :CALC2:PWAV:STAT is on, the power-
weighted average wave number is returned.
Attribute Summary
Preset State: not affected
SCPI Compliance: standard
Query Only
Description
Use the CALC2:POIN? query to determine the number of points the
CALC2:DATA? query will return. The following string is a typical example of the
first few returned values returned when WAVelength is specified:
+1.54488600E-006,+1.54649100E-006,+1.54808300E-006,+1.54969600E-
006,+1.55131200E-006,+1.55293000E-006
This next string resulted by specifying the WNUMber argument:
+6.47005,+6.46005,+6.45005,+6.45005,+6.446155
00E+005,+6.43005
Notice that only measurement values are returned to the computer. There is no first
value that indicates the number of values contained in the string as there is, for
example, with the FETCh, READ, and MEASure commands.
When there is no input signal, the POWer query returns –200 dBm; the WAVelength
query returns 100 nm (1.0E–7).
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......