![Keysight Technologies 86120B Скачать руководство пользователя страница 180](http://html1.mh-extra.com/html/keysight-technologies/86120b/86120b_user-manual_1961930180.webp)
178
Programming Commands
CALCulate3 Subsystem
DATA?
Queries the data resulting from delta, drift, and signal-to-noise measurements.
Syntax
:CALCulate3:DATA? {POWer | FREQuency | WAVelength | WNUMber}
Argument
Description
POWer
Queries the array of laser-line powers after the calculation is
completed.
FREQuency
Queries the array of laser-line frequencies after the calculation
is completed.
WAVelength
Queries the array of laser-line wavelengths after the calculation
is completed.
WNUMber
Queries the array of laser-line wave numbers after the
calculation is completed.
Attribute Summary
Preset State: not affected
SCPI Compliance: standard
Query Only
Description
The data returned by the query depends upon which calculation state is on. If no cal-
culation state is on, an error is generated. The returned data is comma delimited. The
following string is a typical example of six values returned when POWer is specified
from a delta power measurement:
-7.42000,-1.00000,-2.52000,
-3.41000,-3.80000,-6.36000
Notice that only measurement values are returned to the computer. There is no first
value that indicates the number of values contained in the string as there is, for
example, with the FETCh, READ, and MEASure commands. Use the
CALC3:POIN? query to determine the number of points the CALC3:DATA? query
returns.
In the SNR or ASNR calculation, only the POWer argument is valid. The other argu-
ments will generate a “Settings conflict” error. Use the CALC2:DATA? query to
retrieve the signal wavelengths and powers.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......