![Keysight Technologies 86120B Скачать руководство пользователя страница 152](http://html1.mh-extra.com/html/keysight-technologies/86120b/86120b_user-manual_1961930152.webp)
150
Programming Commands
Measurement Instructions
MEASure{:ARRay | [:SCALar]} :POWer:FREQuency?
Returns frequency values.
Syntax
:POWer:FREQuency? [<expected_value>[,<resolution>]]
Description
When used with a :SCALar command, a single value is returned. The display is
placed in the single-wavelength mode, and the marker is placed on the signal having
a frequency that is closest to the
<expected_value>
parameter. Default units for
<expected_value>
parameter are in Hz. When used with an :ARRay command, an
array of frequencies is returned. The display is placed in the list-by-wavelength
mode.
The
<resolution>
parameter sets the resolution of the measurement. It is a unitless
number whose value will be limited to either 0.01 or 0.001 (whichever is closer).
MAXimum resolution is equivalent to the FAST measurement update mode. MINi-
mum resolution is equivalent to the NORMAL measurement update mode.
Returned values are in Hz (display is in THz). Power units are not affected.
<expected_value>
Constants
MAXimum
The highest frequency signal
MINimum
The lowest frequency signal
DEFault
The current marker position
Used With
<expected_value>
<resolution>
SCALar
optional
optional
ARRay
ignored
a
a. Although ignored, this argument must be present if the resolution argument is specified.
optional
CONFigure command
When this function is used with the CONFigure command, the query question
mark character “?”
must
not
be included in the string. However, the FETCh,
READ, and MEASure command are queries and require the question mark.
Refer to the examples for this command.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......