![Keysight Technologies 86120B Скачать руководство пользователя страница 224](http://html1.mh-extra.com/html/keysight-technologies/86120b/86120b_user-manual_1961930224.webp)
222
Performance Tests
Test 2. Sensitivity
Test 2. Sensitivity
Description
Sensitivity is verified using the following devices:
• Optical power meter
• Optical attenuator
• 1310 nm and 1550 nm lasers (>0 dBm output power)
C A U T I O N
Do not 18 dBm source power. The Keysight 86120B’s input
circuitry can be damaged when total input power exceeds 18 dBm.
Procedure
Perform the following procedure first using the 1310 nm laser and then repeat the
steps using the 1550 nm laser.
1
Connect the laser’s output to the optical attenuator’s input.
2
Connect the optical attenuator’s output to the optical power meter.
3
Adjust the attenuator for a reading of 0 dBm on the power meter. Record the
attenuator’s setting.
Attenuation at 0 dBm: ____________
4
Adjust the attenuator for a reading of –35 dBm on the power meter. Record the
attenuator’s setting.
Attenuation at –35 dBm: ____________
5
Disconnect the fiber-optic cable at the power meter’s input and connect the cable to
the Keysight 86120B being tested.
6
Reset the optical attenuator to the setting recorded in
7
Read the power and wavelength measured on the Keysight 86120B, and compared
them to the specifications listed in
Chapter 7, “Specifications and Regulatory
.
8
Reset the optical attenuator to the setting recorded in
9
Read the power and wavelength measured on the Keysight 86120B, and compared
them to the specifications listed in
Chapter 7, “Specifications and Regulatory
.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......