![Keysight Technologies 86120B Скачать руководство пользователя страница 236](http://html1.mh-extra.com/html/keysight-technologies/86120b/86120b_user-manual_1961930236.webp)
234
Specifications and Regulatory Information
Definition of Terms
Flatness
refers to the maximum amplitude error in a measurement between two
lines that are separated in wavelength by no more than the specified amount.
Linearity
indicates the maximum power error in measuring the change in power of
one laser line.
Polarization Dependence
indicates the maximum displayed power variation as the
polarization of the input signal is varied.
Display Resolution
indicates the minimum incremental change in displayed power.
Sensitivity
Sensitivity
is defined as the minimum power level of a single laser-line input to mea-
sure wavelength and power accurately. A laser line with less than the minimum
power may be measured but with reduced wavelength and power accuracy. For mul-
tiple laser-lines input, sensitivity may be limited by total input power.
Selectivity
Selectivity
indicates the ability to measure the wavelength and power of a weak laser
line in the proximity of a specified stronger laser line and separated by the specified
amount.
Input Power
Maximum displayed level
indicates the maximum total input power (total of all laser
lines present) to accurately measure wavelength and power.
Maximum safe input power
indicates the maximum total input power (total of all
laser lines present) to avoid permanent optical damage to the instrument.
Maximum Number
of Lines Input
Maximum number of lines input
is the maximum number of displayed lines. If more
than 100 lines are input, only the 100 longest wavelength lines are displayed.
Input Return Loss
Input Return Loss
indicates the optical power reflected back to the user’s fiber cable
relative to the input power. It is limited by the return loss of the front-panel connec-
tor, and assumes the user’s connector is good.
Measurement Cycle
Time
Measurement cycle time
refers to the cycle time when measuring wavelength and
power of laser lines. Specific advanced applications may require longer cycle times.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......