Performance Tests
Test 1. Absolute Wavelength Accuracy
Test 1. Absolute Wavelength Accuracy
Description
Wavelength accuracy is verified using traceable light sources such as the following
devices:
• Stable lasers
• Gas lamps
• HeNe gas lasers
C A U T I O N
Do not 18 dBm source power. The Keysight 86120B’s input
circuitry can be damaged when total input power exceeds 18 dBm.
Procedure
Use three or four light standards that cover the Keysight 86120B’s wavelength
range. Connect the traceable sources to the Keysight 86120B and verify that the
Keysight 86120B is reading the sources to within the absolute wavelength accuracy
specification.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......