166
Programming Commands
CALCulate2 Subsystem
Description
A laser line is identified as a valid peak if its amplitude is above the maximum
amplitude minus the peak threshold value. The subtraction is done in dB units. This
setting works in conjunction with the peak excursion setting to determine which
responses are located.
Refer to “PEXCursion” on page 164
. Changing the peak
threshold limit causes the instrument to reprocess the current set of data. Refer also
to
“To define laser-line peaks” on page 54
The query response is the current value. For example, if the current value is set to 15
dB, the following value is returned:
15
PWAVerage[:STATe]
Places the instrument in the power-weighted average mode.
Syntax
:CALCulate2:PWAVerage[:STATe]{?| {ON | OFF | 1 | 0}}
Attribute Summary
Preset State: off
*RST State: off
Description
When the state is on, the CALC2:DATA? POW query returns the total power and
the CALC2:DATA? WAV, FREQ, or WNUM query returns the power-weighted
average wavelength, frequency, or wave number values.
Turning power-weighted average mode on while making delta, coherence length, or
signal-to-noise measurements results in a “–221 Settings conflict” error.
WLIMit[:STATe]
Limits input wavelength range of the Keysight 86120B.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command
has the time to complete before sending any more commands to the instrument.
Refer to “Always force the Keysight 86120B to wait for non-sequential com-
mands” on page 92
for more information.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......