![Keysight Technologies 86120B Скачать руководство пользователя страница 97](http://html1.mh-extra.com/html/keysight-technologies/86120b/86120b_user-manual_1961930097.webp)
95
Programming
Making Measurements
The format of returned data
Measurements are returned as strings
All measurement values are returned from the Keysight 86120B as ASCII strings.
When an array is returned, the individual values are separated by the comma charac-
ter.
Determine the number of data points
When a FETCh, READ, or MEASure command is used (with ARRay specified), the
first returned value indicates the total number of measurement values returned in the
query.
If you use the:CALCulate1:DATA?, :CALCulate2:DATA?, or :CALCulate3:DATA?
queries to query data, send the :POINts? query first to determine the number of val-
ues returned in the string. The string does not contain a first value which specifies
the string length. This is shown in the following example:
OUTPUT 720;”:CALCulate1:POINts?”
ENTER 720;Length
OUTPUT 720;”:CALCulate1:DATA?”
ENTER 720;Result$
Data can be corrected for elevation and vacuum
Normally, the Keysight 86120B provides measurement values calculated for condi-
tions in air at sea level. Use the :SENSe:CORRection:ELEVation command to com-
pensate for air dispersion. Altitudes up to 5000 meters can be entered. Use the
:SENSe:CORRection:MEDium command to switch to readings in a vacuum.
Amplitude units
The default amplitude units are dBm. If you need measurements in watts, use the
:UNIT:POWer command. When the Keysight 86120B is turned on, the amplitude
units are automatically set to the units used before the instrument was last turned off.
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......