Using the Multi-Wavelength Meter
Displaying Wavelength and Power
To display average wavelength and total power
• Press the Avg WL key.
Measuring lasers between 700 nm and 1200 nm
After the Preset key is pressed, the input wavelength range is limited to measuring
lasers between 1200 nm and 1650 nm. This prevents the accidental display of spuri-
ous signals that may not exist. You can easily expand the input range to the full 700
nm to 1650 nm range, however you should learn how to identify spurious signals.
Spurious signals below 1200 nm may be displayed whenever low-power laser lines
(power levels near the Agilent 86120B’s specified sensitivity) are present at the
input. For example, a low-power laser line at 1550 nm has a second harmonic line at
775 nm. If this second harmonic is above the peak threshold level relative to the fun-
damental line, it is considered a peak. Its displayed power level may be
greater
than
that of the fundamental because the amplitude correction at 775 nm is much greater
(by about 15 dB) than that at 1550 nm (the interferometer is less sensitive at 775
nm).
You can also avoid displaying this second harmonic line by reducing the peak
threshold below its preset value. Because the peak threshold level is used to deter-
mine which signals are to be displayed
before
amplitude corrections are applied, the
harmonic will be eliminated.
Refer to “Defining Laser-Line Peaks” on page 52
.
To use the full wavelength range
1
Press the Setup key.
2
Press the
WL LIM
softkey.
3
Press
LIM OFF
to remove the limits on wavelength range. All responses in the full
700 nm to 1650 nm range are now displayed.
Limiting the wavelength range
The wavelength range of measurement can be limited with the wavelength limit
function. Both start and stop wavelengths can be chosen. The units of wavelength
start and stop are the same as the currently selected wavelength units. If wavelength
Содержание 86120B
Страница 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Страница 2: ......
Страница 3: ...1...
Страница 12: ...10 General Safety Considerations...
Страница 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Страница 100: ...98 Programming Monitoring the Instrument...
Страница 163: ...161 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268...
Страница 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Страница 232: ......
Страница 233: ...7 Definition of Terms 233 Specifications 235 General Safety Information 239 Specifications and Regulatory Information...
Страница 244: ...242 Specifications and Regulatory Information Product Overview...
Страница 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Страница 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Страница 253: ...251 Reference Menu Maps System Print Menu...
Страница 254: ...252 Reference Menu Maps System Setup Menu...
Страница 272: ......